top of page
DIN Rail Mounted Temperature Transmitter – Thermocouple & RTD

DIN Rail Mounted Temperature Transmitter – Thermocouple & RTD

From $135.00Sale Price
Input/Output
1 in/ 1out
2 in/ 2 out
Input Type
RTD (PT100)
Thermocouple
PT1000
Output
4-20mA
0-5V
0-10V
RS485
0/500
Quantity

DIN rail temperature transmitter for thermocouple and RTD sensors. Programmable inputs, isolated signal output (4–20mA, 0–20mA, 0–5V, 1–5V), high accuracy, and compact design for PLC, DCS, and control systems.

Conforms to SIL 2 systems

  • The DIN Rail Mounted Temperature Transmitter is a compact, high-accuracy device designed to convert thermocouple and RTD sensor signals into standard analog outputs for industrial control systems. With programmable input types, isolated signal transmission, and multiple output options, it is ideal for integrating temperature measurement into PLCs, DCS, and instrumentation networks.

    This transmitter supports 1 input 1 output, 1 input 2 outputs, or 2 inputs 2 outputs configurations, offering flexibility for different applications. Featuring a slim DIN35mm rail mount design with pluggable terminals, it allows fast installation in control panels. Isolation between power, input, and output ensures reliable operation in harsh industrial environments.

    Key Features

    • Programmable for thermocouple and RTD inputs

    • Configurations: 1 input 1 output, 1 input 2 outputs, or 2 inputs 2 outputs

    • Transmission accuracy (at 20°C): (0.2%FS + 1 digit)

    • Temperature drift: 0.0075%FS/°C

    • Galvanic isolation: 1.5kVrms, 1 min, 50Hz (Power ↔ Input ↔ Output)

    • Standard output signals: 4–20mA, 0–20mA, 0–5VDC, 1–5VDC

    • DIN35mm rail mounting with pluggable terminals

    • Slim housing: 12 × 108 × 113 mm

    • Power supply: 20–32VD

    Typical Applications

    • Industrial automation and control systems

    • PLC and DCS temperature signal integration

    • Process monitoring and instrumentation

    • Energy, HVAC, and building management systems

  • Input type Thermocouple, RTD Input Type Measured Range
    . Output Signal 4-20mA, 0-10mA, 0-20mA; 0-5VDC, 1-5VDC RTD Pt100 -200 to 800℃
    Resistor Load (mA output) RL≤400Ω (4-20ma, 0-10ma, 0-20ma) CU50 -50 to150℃
    Resistor Load (Vdc output) RL≥250kΩ (1-5VDC, 0-5VDC) CU100 -50 to150℃
    . Accuracy 0.2FS%+1 digit CU53* -50 to 150.0℃
    Temperature shift 0.0075%FS/℃ BA1 -200 to 600.0℃
    Min. Resolution 0.1℃ BA2 -200 to 600.0℃
    Cold junction -compensation -10 to 50℃ T.C K -200 to 1300℃
    Cold junction-comp. accuracy ±1℃ T -200 to 400℃
    Response time 50ms up to 90% of final value E -200 to 1000℃
    Isolated ability Input-output-power 1.5KV, 1min, 50Hz J 0 to 1200℃
    Insulation resistance Input-output-power≥100mΩ at 500VDC N 0 to 1300℃
    Power supply DC20-32V R 0 to 1600℃
    Power Consumption ≤1W (1 input 1Output); ≤1.4W(1/2Input2Output)  S 0 to 1600℃
    Case material/Protection PC/ IP20 B 400 to 1800℃
    Size 121.2Xx108x13mm Wre3-25 0-2300℃
    Working T&H -10 to 50℃; 25% to 85%RH Wre5-26 0-2300℃
    Storage temperature -10 to 60℃ (No dew) F2* 700-2000℃
bottom of page